MENU
RADI PCR 96 is based on global vision of product design concepts and manufacturing processes.
It creatively combines Fresnel lens optical signal acquisition technology, time-resolved signal separation technology and unique temperature control technology.
And it reaches international advanced level in sensitivity, multi-color crosstalk, temperature uniformity and accuracy.
It supports the application of all common QPCR detection modes.
· Up to 6 fluorescence detection channels allowing multiplex PCR- Simultaneous detection of 5 target genes in 96 samples
· Innovative scanning method and time-resolved signal separation technology - High accuracy
· Unique edge temperature compensation technology
TEMPERATURE CONTROL SYSTEM | |
---|---|
Sample capacity |
0.1ml PCR tubes x 96, 8 x 12 PCR plate or 96 well plate x 1 |
reaction volume |
10-50 μl |
thermal cycle technology |
7-inch Touch LCD screen |
electrical specification |
Peltier |
Max. Heating/ Cooling rate |
6.0℃/s |
Heating temperature range |
4 - 100 ℃ |
Temperature accuracy |
± 0.2℃ |
Temperature uniformity |
± 0.2℃ @60℃, ± 0.2℃ @95℃ |
Temperature gradient setting range |
30 ℃ ~ 100 ℃ |
Temperature gradient difference setting range |
1 - 36℃ |
Detection system | |
---|---|
excitation light source |
4/6 monochrome high efficiency LEDs |
Detection device |
PMT |
Detection mode |
Time-resolved signal separating technology |
Excitation/ detection wavelength range |
455-650nm / 510-715nm |
Fluorescent channels |
4/6 channels |
supported dye |
FAM/SYBR Green, VIC/JOE/HEX/TET, ABY/NED/TAMRA/Cy3, JUN, ROX/Texas Red, Mustang Purple, Cy5/LIZ |
sensitivity | Single copy gene |
Resolution |
1.33 folds copy number difference can be distinguished in single-plex qPCR |
Dynamic range | 10 orders of magnitude copies |